8Firmware Update - R&S ESU Series Operating Manual

Emi test receiver
Table of Contents

Advertisement

R&S ESU
In case of failure a short description of the failed test, the defective module, the
associated value range and the corresponding test results are indicated.
Remote command:
PAGE UP /
The PAGE UP or PAGE DOWN softkey sets the SELFTEST RESULTS table to the
PAGE DOWN
next or previous page.
Remote command:
Hardware Adjustment
Some of the R&S ESU modules can be realigned. This realignment can become
necessary after calibration due to temperature drift or aging of components (see ser-
vice manual instrument).
The realignment should be carried out by qualified personnel since the changes
considerably influence the measurement accuracy of the instrument. This is the
reason why the softkeys REF FREQUENCY, CAL SIGNAL POWER and SAVE
CHANGES can only be accessed after entering a password.
4.9.3.8
Firmware Update
For the latest firmware version refer to the Rohde & Schwarz internet site and down-
load the most up-to-date firmware.
A detailed description about performing the firmware update in the intrument is given
in the release note, provided on this CD. The release note is also downloadable
from the Rohde & Schwarz internet site.
The installation of a new firmware version can be performed using a memory stick.
The installation program is called in the SETUP menu.
FIRMWARE
FIRMWARE UPDATE
UPDATE
RESTORE FIRMWARE
UPDATE PATH
4.296
DIAG:SERV:STE:RES?
--
Operating Manual 1302.6163.12 - 03
Instrument Functions
Basic Settings

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents