R&S ESU Series Operating Manual page 486

Emi test receiver
Table of Contents

Advertisement

R&S ESU
CALCulate<1|2>:MARKer<1...4>:Y:PERCent 0 to100%
This command positions the selected marker in the selected window to the given probability. If marker
2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
The command is only available with the CCDF measurement switched on.
The associated level value can be determined with the CALC:MARK:X? command.
Example:
"CALC1:MARK:Y:PERC 95PCT"
'Positions marker 1 in screen A to a 'probability of 95%.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:MAXimum[:PEAK]
This command positions the specified marker to the current maximum value of the corresponding
trace. The corresponding marker is activated first or switched to the marker mode.
If no maximum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error code: -200) is produced.
Example:
"CALC:MARK2:MAX"
'Positions marker 2 in screen A to the maximum value of 'the trace.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MAXimum:NEXT
This command positions the specified marker to the next smaller maximum value of the corresponding
trace.
If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.
Example:
"CALC:MARK2:MAX:NEXT"
'Positions marker 2 in screen A to the next 'lower maximum value.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.47