R&S ESU Series Operating Manual page 251

Emi test receiver
Table of Contents

Advertisement

R&S ESU
LIST
The LIST EVALUATION softkey activates or deactivates the spurious emission mea-
EVALUATION
surement LIST EVALUATION. The peak search evaluation is automatically done
during measurement and the results are displayed in a table in the lower part of the
screen.
Following results are listed:
If LIST EVALUATION is switched on, the functions PEAKS PER RANGE, MARGIN,
PEAK SEARCH and VIEW PEAK LIST are not available.
Remote command:
IF SHIFT
Refer to the
STOP MEAS
See
START MEAS
See
PEAK SEARCH
The PEAK SEARCH softkey starts the process of determining the list of the sub-
range maximums from the existing sweep results. This procedure can be repeated
as often as desired, e.g. to experiment with the various threshold settings.
It is activated only after a measurement has been performed with START MEAS.
Remote command:
PEAKS PER
The PEAKS PER RANGE softkey activates entry of the number of peaks per range
RANGE
that are stored in the list. The value range extends from 1 to 50. Once the selected
number of peaks has been reached, the peak search is stopped in the current range
and continued in the next range. The default value is 25.
Remote command:
4.220
frequency range
frequency of the absolute peak power in this range
level delta to the limit with margin in dB
limit check state (by color change and an asterisk at the end of the line if failed)
IF SHIFT
softkey in the setup menu.
"STOP MEAS" on page
"START MEAS" on page
CALC1:PEAK:AUTO ON | OFF
4.219.
4.219.
CALC:PEAK
CALC:PEAK:SUBR 1...50
Operating Manual 1302.6163.12 - 03