R&S ESU
SYSTem:PASSword[:CENable] 'password'
This command enables access to the service functions by means of the password.
Example:
"SYST:PASS 'XXXX'"
Characteristics: *RST value: –
SCPI: conform
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
SYSTem:PRESet
This command initiates an instrument reset.
The effect of this command corresponds to that of the PRESET key with manual operation or to the
*RST command.
Example:
"SYST:PRES"
Characteristics: *RST value: –
SCPI: conform
Mode:
R, A
SYSTem:PRESet:COMPatible FSP | OFF
This command specifies the preset compatibility.
Parameter:
OFF: Mixed mode is set as preset mode.
FSP: Analyzer mode is set as preset mode.
Example:
"SYST:PRES:COMP FSP"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
SYSTem:SHUTdown
This command shuts down the instrument.
This command is an event and therefore has no query and no *RST value.
Example:
"SYST:SHUT"
Characteristics: *RST value: -
SCPI: device- specific
Mode:
R, A
SYSTem:SPEaker:VOLume 0 to 1
This command sets the volume of the built-in loudspeaker for demodulated signals. Minimum volume
is set by 0 and maximum volume by 1.
The value 0 is the lowest volume, the value 1 the highest volume.
Example:
"SYST:SPE:VOL
6.244
Remote Control – Description of Commands
0.5"
Operating Manual 1302.6163.12 - 03
SYSTem Subsystem