R&S ESU Series Operating Manual page 478

Emi test receiver
Table of Contents

Advertisement

R&S ESU
6.5.3.4
CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem defines the upper limit line.
CALCulate<1|2>:LIMit<1...8>:UPPer[:DATA] <numeric_value>,<numeric_value>...
This command defines the values for the upper limit lines.
Up to 8 limit lines can be defined at the same time. This command is independent of the measurement
window.
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer limit line
have to be identical. Otherwise default values are entered for missing values or not necessary values
are deleted.
The unit must be identical with the unit selected by CALC:LIM:UNIT. If no unit is indicated, the unit
defined with CALC:LIM:UNIT is automatically used.
In analyzer mode, the limit check signals errors if the measured values exceed the UPPer limit line.
Example:
"CALC:LIM2:UPP -10,0,0,-10,-5"
'Defines 5 upper limit values for limit line 2 in the preset unit.
"CALC:LIM2:UPP?"
'Outputs the upper limit values for limit line 2 separated by a comma.
Characteristics: *RST value: - (LIMit:STATe is set to OFF)
SCPI: conform
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:UPPer:STATe ON | OFF
This command switches on or off the indicated limit line. The limit check is activated separately with
CALC:LIM:STAT ON.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window.
In analyzer mode, the result of the l imit check can be queried with CALCulate:LIMit<1...8>:
FAIL?.
Example:
"CALC1:LIM4:UPP:STAT ON"
'Switches on limit line 4 (upper limit) in screen A.
"CALC2:LIM4:UPP:STAT ON"
'Switches on limit line 4 (upper limit) in screen B.
Characteristics: *RST value: OFF
SCPI: conform
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.39