4Display Of Measurement Results - R&S ESU Series Operating Manual

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Once measurement has been completed, the display of the measured values can
be enlarged down to the individual measured values by using the zoom function.
Thus, each individual click interferer can be evaluated in detail if necessary.
d B µ V
1 P K
C L R W R
2 Q P
C L R W R
Fig. 4.24 Scan in the time range
Time domain analysis is started with the RUN softkey. It can be performed as a sin-
gle operation (SINGLE) or as a repeating operation (CONTINUOUS). In SINGLE
operation, time domain analysis stops after the total run time has elapsed. In CON-
TINUOUS operation, time domain analysis can either be interrupted with the HOLD
SCAN softkey or stopped with the STOP SCAN softkey. The measurement itself is
continuous, i.e. if the end of the graph is reached and recording starts again at the
beginning of the graph, the measurement keeps running internally without any inter-
ruption.
4.5.5.4
Display of Measurement Results
Full-screen or split-screen mode can be selected for result display. In the full-screen
mode, the result diagram covers the whole screen area.
In split-screen mode, the upper half of the screen displays the frequency and level
readout, i.e. a bar graph. The diagram is displayed in the lower half.
Up to 3 detectors can be measured simultaneously. They are assigned to traces 1 to
3. Since the detectors are set only once, it is not possible to measure with different
detectors in different the subranges.
4.54
F r e q 9 3 M H z
A t t 1 0 d B
1 0 0
9 0
8 0
7 0
6 0
T R G 5 4 . 1 d B µ V
5 0
4 0
3 0
2 0
1 0
0
0 s
Instrument Functions
R B W
1 2 0 k H z
M T
1 m s
P R E A M P O F F
1 s /
Operating Manual 1302.6163.12 - 03
Receiver Mode
S G L
T R G
1 0 s