R&S ESU Series Operating Manual page 597

Emi test receiver
Table of Contents

Advertisement

R&S ESU
If LINear is selected, the measurement with logarithmic level scaling yields a much "flatter" falling edge
than LOGarithmic. This behavior is due to the conversion of linear power into logarithmic level. If the
linear power is halved, the level decreases by only 3 dB.
Example:
"BAND:VID:TYPE LIN"
'Video filter ahead of the logarithmic amplifier
Characteristics: *RST value: LIN
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]BANDwidth|BWIDth:PLL AUTO | HIGH | MEDium | LOW
This command defines the bandwidth of the main PLL of the instrument synthesizer.
Example:
"BAND:PLL HIGH"
Characteristics: *RST value: AUTO
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]BANDwidth|BWIDth:IF
This command sets the resolution bandwidth of the IF analysis.
Digitally implemented resolution filters from 10 Hz to 100 kHz in steps of 1, 3, and 10 are available
(each with 6 dB bandwidths). The reliable value range depends on the selected frequency span of the
IF analysis.
Example:
"BAND:IF 10kHz"'Sets the IF bandwidth to 10 kHz.
Characteristics: *RST value:3 kHz
SCPI:device-specific
Mode:
R
6.158
Remote Control – Description of Commands
10 Hz...100 kHz