Sets of Angles
Field
Option
Tol Exceed
Only
Never
<Time Out:>
No Time Out There is no automatic closure of message dialogs.
1 sec to
60 sec
<Timer Monit.:>
Yes
No
Description
Only the message dialog relating to the exceeding of
tolerances is displayed and is closed as defined by the
settings in <Time Out:>.
No message dialogs are displayed except for
specific warnings.
Specific warnings which affect the instrument and
it's ability to continue with the monitoring process
will be displayed and will remain on the screen.
These include the overheating of the instrument,
low battery levels, unavailable space on the
CompactFlash card.
To define the time delay for the automatic closing of
message dialogs during a measurement set. This
choicelist is not available when <Stop For: Never>.
When a message dialog appears, it is only closed by
pressing YES (F4).
All message dialogs are automatically closed as
defined by these individual time settings.
This input field is only available when Monitoring is
registered through the licence key.
Automatic monitoring of target points is activated.
Automatic monitoring of target points is not activated.
The Sets of Angles application will apply.