E5071C
User Defined Variable
The E5071C is having an area in which a User can set any value. These
area are divided by the different data format of the values. A maximum of
ten (1 to 10) such areas can be used by each command.
For example, after setting the value (data) obtained using VBA of the
E5071C to the User defined variable, this value is available to an external
controller (program) through the use of these User defined variables.
Turning E5071C power ON/OFF initializes the User defined
variables. They are not initialized through executing Preset.
These commands does not refers to or change the results of the
E5071C.
SCPI.PROGram.VARiable.ARRay(Vnum).DATA
SCPI.PROGram.VARiable.ARRay(Vnum).SIZE
SCPI.PROGram.VARiable.DOUBle(Vnum).DATA
SCPI.PROGram.VARiable.LONG(Vnum).DATA
SCPI.PROGram.VARiable.STRing(Vnum).DATA