E5071C
For channel 1, turn on the continuous trigger startup mode to On and set
the trigger source to the bus trigger.
Lines 320 to 360
For channel 1, set the sweep center value to the Center variable, the
sweep span value to the Span variable, the number of measurement points
to the Nop variable, the IF bandwidth to the IfBw variable, and the power
level to the Pow variable.
Lines 380 to 410
For channel 1, set the number of traces to the NumTrac variable, the
measurement parameter to the Par variable, and the data format to the
Fmt variable.
Line 450
Stores the calibration kit number for channel 1 into the CalKit variable.
Line 460
Stores 1 and 2 into the Port variable that indicates ports used for the full
2-port calibration.
Line 480
Calls the Calib_Solt procedure (lines 1200 to 2130).
Lines 520 to 530
Save the instrument setting and the calibration coefficient into a file whose
name is specified with the File variable.
Line 580
Displays a message that prompts you to connect a DUT (Device Under
Test) in the instrument status bar in the lower part of the LCD display and
waits for the operation of
> Continue after the connection.
Macro Setup
Lines 620 to 630
Generate a trigger to start a single sweep and wait until the measurement
finishes (1 is read out with the SCPI.IEEE4882.OPC object).
Line 650
For trace 1 of channel 1, executes auto scale to set the optimum scale.
Lines 690 to 710
Display marker 1 and move it so that the stimulus value becomes equal to
the value of the Center variable. Then, these lines read out the response
value of marker 1 and store it into the MkrVal variable.
Line 730