Making Measurements
Using Test Sequencing to Test a Device
To run the sequence, press:
SEQUENCE 1 SEQ 1
Preset

Limit Test Example Sequence

This measurement example shows you how to create a sequence that will branch the
sequence according to the outcome of a limit test. Refer to
Device" on page
1-75, for a procedure that shows you how to create a limit test.
For this example, you must have already saved the following in register 1:
• device measurement parameters
• a series of active (visible) limit lines
• an active limit test
1. To create a sequence that will recall the desired instrument state, perform a limit test,
and branch to another sequence position based on the outcome of that limit test, press:
NEW SEQ MODIFY

Seq

RECALL KEYS MENU
Save/Recall
SPECIAL FUNCTIONS
Seq
IF LIMIT TEST PASS
IF LIMIT TEST FAIL
Seq
DONE SEQ MODIFY
This will create a displayed list for sequence 1, as shown:
Start of Sequence
RECALL REG 1
IF LIMIT TEST PASS THEN DO
SEQUENCE 2
IF LIMIT TEST FAIL THEN DO
SEQUENCE 3
2. To create a sequence that stores the measurement data for a device that has passed the
limit test, press:
NEW SEQ MODIFY
Seq
SELECT DISK
Save/Recall
DEFINE DISK-SAVE
Seq
DONE SEQ MODIFY
This will create a displayed list for sequence 2, as shown:
Start of Sequence
INTERNAL DISK
DATA ARRAY
1-104
SEQUENCE 1 SEQ1
RECALL REG1
DECISION MAKING
SEQUENCE 2 SEQ2
SEQUENCE 3 SEQ3
SEQUENCE 2 SEQ2
INTERNAL DISK
DATA ARRAY ON
RETURN
"Using Limit Lines to Test a
RETURN
SAVE STATE