Agilent Technologies E5071C Manual page 1176

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Programming
Lines 450 to 460
For traces 1 and 2, these lines executes the auto scale to set the optimum
scale.
Lines 490 to 500
Read out the value of the questionable limit status event register and store
the AND of the read-out value and 2 (the value in which only bit 1 is 1)
into the Test_Ch1 variable.
Lines 510 to 530
Read out the value of the questionable limit channel 1 status event register
and store the AND of the read-out value and 2 (the value in which only bit
1 is 1) into the Test_Tr1 variable and the AND of the read-out value and 4
(the value in which only bit 2 is 1) into the Test_Tr2 variable.
Lines 560 to 610
Specify trace 1 as the active trace and set the analysis range (start point:
1.85 GHz and stop point: 1.91 GHz). Then these lines sets the polarity of
the peak search (both the positive peak and the negative peak) and the
lower limit of the peak excursion value (0).
Lines 620 to 640
Search for the minimum value within the analysis range and store the
analysis result into the IlossTx variables.
Line 650
Uses the ripple analysis function to store the maximum value of the
response differences between the positive peaks and the negative peaks
within the analysis range into the RipTx variables.
Lines 670 to 720
Specify trace 2 as the active trace and set the analysis range (start point:
1.93 GHz and stop point: 1.99 GHz). Then these lines set the polarity of
the peak search (both the positive peak and the negative peak) and the
lower limit of the peak excursion value (0).
Lines 730 to 750
Search for the minimum value within the analysis range and store the
analysis result into the IlossRx variables.
Line 760
Uses the ripple analysis function to store the maximum value of the
response differences between the positive peaks and the negative peaks
within the analysis range into the RipRx variables.
Line 790
1403