is used to parse the messages of SCPI commands. For instructions on
using the E5071C's VBA Editor to create a program that uses COM objects,
refer to Operation Basics of the E5071C's VBA.
Major Control Difference between COM Object and SCPI Command
While the control using SCPI commands allows SRQ (Service Request)
interrupts through the status reporting mechanism, the control using COM
objects does not support SRQ interrupts. Instead of SRQ interrupts, you
can use the
object to suspend the program until the E5071C is
WaitOnSRQ
put into the desired state.
1314