Multi Test Voltage Measurement Sample Program - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B
Figure 6-4. Multi Test Voltage Measurement Sample Program
Lines 70 and 80
Line 100
Lines 110 and 120
Line 130
Lines 140 to 160
Lines 180 to 260
Lines 280 and 290
Line 300
Line 310
Line 320
Line 330
Line 340
Lines 360 to 380
Lines 400 and 420
Line 440
Lines 460 to 490
Lines 530 to 600
Lines 620 to 640
Resets the 4349B and clears the status byte register.
Selects the resistance measurement mode.
Enters the voltage value of 1 V for channel 1, and 100 V for channel
2.
Sets the measurement time mode to LONG.
Sets the measurement range to 100 pA for channel 1 and channel 2.
Performs an OPEN correction, and prompts the operator to connect
the DUT.
Sets the comparator limit values for channel 1 and channel 2; the
2
lower limit to 1
10
12
Enables the comparator function.
Clears the LCD display.
Enables the contact check function.
Sets the trigger source mode to External.
Sets the data transfer format to the 64-bit real transfer format.
Sets the size of the data bu er to 50, sets up to feed the measurement
data to the data bu er, and sets up to feed data into the data bu ers
whenever new data is available.
Enables the DBUF full bits of the operation status register, and the
operation status register summary bit of the status byte register.
Triggers the measurement from an external trigger source.
Waits until the data bu ers are full.
Retrieves the measurement data held in DBUF.
Prints the measurement data.
and the upper limit to the 1
2
10
ohm;.
13
Application Measurement
6-7