Line 1490
Sets channel 1's segment table to the SegmData variable.
Line 1500
Sets channel 1's sweep type to "segment".
Line 1510
Sets the channel 1 graph's horizontal axis display method to the order
base (the axis on which the measurement point numbers are placed evenly
in the order of measurement).
Procedure: Calib_Solt (lines 1550 to 2420)
See Lines 1200 to 2130 of Calib_Solt.
Procedure: Setup_Limitline (lines 2440 to 3180)
Line 2490
Stores the number of lines (5) in trace 1 limit table into the LimDataS13(0)
variable.
Lines 2500 to 2790
Store the settings in trace 1 limit table into the LimDataS13(1) to
LimDataS13(25) variables.
Line 2820
Stores the number of lines (4) in trace 2 limit table into the LimDataS21(0)
variable.
Lines 2830 to 3060
Store the settings in trace 2 limit table into the LimDataS21(1) to
LimDataS21(20) variables.
Lines 3080 to 3110
Specify trace 1 as the active trace, store trace 1's limit line into the
LimDataS13 variable, and display it. Then, these lines turn on the limit test
function for trace 1.
Lines 3130 to 3160
Specify trace 2 as the active trace, store trace 2's limit line into the
LimDataS21 variable, and display it. Then, these lines turn on the limit test
function for trace 2.
Procedure: Setup_Register (lines 3200 to 3260)
Lines 3220 to 3230
Programming
1405