Turns off display update. Turning off display update shortens drawing time
and object processing time.
Line 140
Returns the E5071C to the preset state.
Lines 170 to 180
For channel 1, these lines turn the continuous trigger startup mode to On
and sets the trigger source to the bus trigger.
Line 200
Calls the Setup_Parameter procedure (lines 910 to 1140). For information
on the Setup_Parameter procedure, see the description later.
Line 220
Calls the Setup_Segment procedure (lines 1160 to 1530). For information
on the Setup_Segment procedure, see the description later.
Line 250
Stores the calibration kit number for channel 1 into the CalKit variable.
Line 260
Stores 1, 2, and 3 into the Port variable that indicates the ports used for
the full 3-port calibration.
Line 280
Calls the Calib_Solt procedure.
Line 310
Calls the Setup_Limitline procedure (lines 2420 to 3180). For information
on the Setup_Limitline procedure, see the description later.
Line 330
Calls the Setup_Register procedure (lines 3200 to 3260). For information
on the Setup_Register procedure, see the description later.
Line 380
Displays the message that prompts for connecting a DUT (Device Under
Test) and waits for the OK button to be clicked after the connection.
Line 410
Clears the questionable limit status event register and questionable limit
channel 1 status event register.
Lines 420 to 430
Generate a trigger to start a single sweep and wait until the measurement
finishes (1 is read out with the SCPI.IEEE4882.OPC object).