Programming
Assigns a GPIB address to the I/O pass.
Lines 80 to 110
These lines set the sweep type to linear sweep, the sweep center value to 950 MHz, the sweep span
value to 100 MHz, and the number of measurement points to 201.
Lines 120 to 160
These lines set the trigger source to bus trigger, turn on Continuous Activation mode for channel 1, and
turn the mode off for channels 2 through 9.
Lines 180 to 190
These lines display the window for channel 1 only and arrange two graphs tiled horizontally.
Lines 210 to 270
These lines set the number of traces for channel 1 to 2, the measurement parameter and its data format
for trace 1 to S21 and Log Mag, respectively, and those for trace 2 to S11 and Log Mag, respectively.
Line 290
This line turns Off the updating of the LCD screen.
Line 300
This line sets the data transfer format to binary.
Lines 320 to 540
These lines repeat the following procedure ten times.
Lines 340 to 360: These lines trigger the instrument and wait until the measurement cycle finishes.
Lines 400 to 440: Reads out the formatted data array of trace 1 in channel 1.
Lines 460 to 500: Reads out the formatted data array of trace 2 in channel 1.
Line 540: This line updates the LCD screen once.