E5071C
390| SCPI.CALCulate(1).PARameter(1).DEFine = Par
400| SCPI.CALCulate(1).PARameter(1).Select
410| SCPI.CALCulate(1).SELected.Format = Fmt
420|
430| '''Performing full 2-port calibration
440|
450| SCPI.SENSe(1).CORRection.COLLect.CKIT.Select = CalKit
460| Port = Array(1, 2)
470|
480| Calib_Solt 1, 2, Port
490|
500| '''Saving state & cal data
510|
520| SCPI.MMEMory.STORe.STYPe = "CST"
530| SCPI.MMEMory.STORe.STATe = File
540|
550| '''Connecting DUT
560|
570| Meas_Start:
580| Prompt ("Connect DUT, and then press [Macro Setup]-Continue button.")
590|
600| '''Performing single sweep
610|
620| SCPI.TRIGger.SEQuence.SINGle
630| Dmy = SCPI.IEEE4882.OPC
640|
650| SCPI.DISPlay.WINDow(1).TRACe(1).Y.SCALe.AUTO
660|
670| '''Analyzing the results
680|
690| SCPI.CALCulate(1).SELected.MARKer(1).STATe = True
700| SCPI.CALCulate(1).SELected.MARKer(1).X = Center
710| MkrVal = SCPI.CALCulate(1).SELected.MARKer(1).Y
720|
730| On Error GoTo Bw_Err
740|
750| SCPI.CALCulate(1).SELected.MARKer(1).BWIDth.THReshold = -3
1394