the calibration data measurement for all standards is complete, the
message "All calibration data completion" is displayed, and the calibration
coefficient is calculated.
When you cancel the calibration data measurement before
completing the measurement of necessary calibration data, the
settings conditions may not return to its original state.
Then, the limit line is set and the setting required for the limit test
judgment is performed.
The message "Set the DUT between test cables." is displayed. Connect the
DUT (duplexer) between the test cables and then click Yes. The limit line is
set and a single sweep is executed. Then, for both trace 1 (Tx: S13) and
trace 2 (Rx: S21), the Pass/Fail judgment of the limit test and the insertion
loss and ripple analysis result within the pass band are displayed.
Click Exit on the user form displaying the analysis result. The message
prompting for remeasurement is displayed. To perform remeasurement,
click Yes, otherwise click No to terminate the program. The details of the
program within the user form for displaying the analysis result (object
name: frmDupRes) are not described here.
Display of the execution result of the program of "apl_sys.vba"
The duplexer measurement program (object name: mdlDupMeas)
is described in detail below. Line numbers are added for
description purpose only and do not appear in the actual program
source code.
Line 90
Stores the calibration kit number (4: 85032F) into the CalKit variable.
Line 110
Programming
1401