Displays 9 channel windows.
Line 140
Sets the trigger source to "bus".
Lines 160 to 240
These lines turn on or off Continuous Activation mode for each channel
depending on whether the corresponding option buttons are on or off. By
default, the mode is turned on for channel 1 only.
Lines 260 to 270
These lines configure the instrument so that operation status event
register's bit 4 is set to 1 only when operation status condition register's
bit 4 is changed from 1 to 0 (negative transition).
Line 280
Enables the operation status event register's bit 4.
Line 290
Enables the status byte register's bit 7.
Line 300
Clears the status byte register and operation status event register.
Line 310
Triggers the instrument to start a measurement cycle.
Line 320
Verifies that the instrument is in a measurement cycle, and suspends the
program until the end of measurement. The time-out is set to 100 seconds
(maximum value).
Lines 330 to 350
These lines display a measurement completion message upon detecting
the end of measurement.
Line 370
Displays the UserForm (object name :frmSrqMeas) on the screen.
Programming
1357