Limit Test Example Sequence - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Limit Test Example Sequence

This measurement example shows you how to create a sequence that will branch the sequence according
to the outcome of a limit test. Refer to
shows you how to create a limit test.
For this example, you must have already saved the following in register 1:
• device measurement parameters
• a series of active (visible) limit lines
• an active limit test
1. To create a sequence that will recall the desired instrument state, perform a limit test, and branch to
another sequence position based on the outcome of that limit test, press:
Seq
NEW SEQ MODIFY
Save/Recall
RECALL KEYS
Seq
SPECIAL FUNCTIONS
IF LIMIT TEST PASS
IF LIMIT TEST FAIL
DONE SEQ MODIFY
Seq
This will create a displayed list for sequence 1, as shown:
Start of Sequence
RECALL REG 1
IF LIMIT TEST PASS THEN DO
SEQUENCE 2
IF LIMIT TEST FAIL THEN DO
SEQUENCE 3
2. To create a sequence that stores the measurement data for a device that has passed the limit test, press:
Seq
NEW SEQ MODIFY
Save/Recall
SELECT DISK
DEFINE DISK-SAVE
Seq
DONE SEQ MODIFY
This will create a displayed list for sequence 2, as shown:
Start of Sequence
INTERNAL DISK
DATA ARRAY
ON
FILENAME
FILE 0
SAVE FILE
"Using Limit Lines to Test a Device" on page
SEQUENCE 1 SEQ1
RECALLKEYS MENU
DECISION MAKING
SEQUENCE 2 SEQ2
SEQUENCE 3 SEQ3
SEQUENCE 2 SEQ2
INTERNAL DISK
DATA ARRAY ON
RETURN
Making Measurements
Using Test Sequencing to Test a Device
1-74, for a procedure that
RECALL REG1
RETURN
SAVE STATE
1- 119

Advertisement

Table of Contents