Lines 1450 to 1520
Display the message that prompts for connecting the open standard to the
specified port. These lines start the measurement of the open calibration
data initiated by clicking OK after the connection and wait for the
completion of the measurement. Click Cancel to return to the beginning of
the calibration.
Lines 1540 to 1610
Display the message that prompts for connecting the short standard to the
specified port. These lines start the measurement of the short calibration
data initiated by clicking OK after the connection and wait for the
completion of the measurement. Click Cancel to return to the beginning of
the calibration.
Lines 1630 to 1700
Display the message that prompts for connecting the load standard to the
specified port. These lines start the measurement of the load calibration
data initiated by clicking OK after the connection and wait for the
completion of the measurement. Click Cancel to return to the beginning of
the calibration.
Lines 1750 to 1840
Display the message that prompts for connecting the thru standard
between the specified ports. These lines start the measurement of the thru
calibration data initiated by clicking OK after the connection and wait for
the completion of the measurement. Click Cancel to return to the beginning
of the calibration.
Lines 1880 to 2060
When the calibration type is not the 1-port calibration (a value other than
1 is specified for the SoltType variable), displays the message asking you
whether you want to measure the isolation calibration data. When Yes is
clicked, displays the message that prompts for connecting the load
standard to the specified two ports (specified with the Port(I-1) and Port(J-
1) variables). These lines start the measurement of the isolation calibration
data initiated by clicking OK after the connection and wait for the
completion of the measurement. Click Cancel to return to the beginning of
the calibration.
Lines 2080 to 2090
Calculate the calibration coefficients from the measured calibration data
and turn on the error correction function. Then, these lines display a
calibration completion message.