DD+DIS150.06E
3.3.2
Safety Note for Replacement of the Power Supply
DANGER:
If the cables at the input side of the power supply are mounted wrong, the
housing of the digitizer can be connected to high voltage. Risk of live.
Mount the cables at the power supply as indicated in the replacement instructions.
3.3.3
Safety Note for Replacement of Cassette Edge Sensor
CAUTION:
The nuts of the cassette edge sensor mounting can be lost very easily.
•
Only loosen the screws of the CASSETTE EDGE SENSOR approx. 2 turns.
•
Always have a "small parts kit" CM+9 5175 9070 0 with you: This contains 2
screws and nuts for the cassette edge sensor mounting.
3.3.4
Safety Note for Replacement of PMT with Light Collector
CAUTION:
The light collector can break when mechanically stressed.
Handle the light collector with great care.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2009
Safety Guidelines / General Repair Instruction
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 /100/110
Repair and Service
Chapter 3.1 / 10
Agfa Company Confidential