AGFA CR 30-X Service Manual page 196

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(14) Open 4 screws at the opener stepper motor
(only 2 screws are shown in Figure 129).
(15) Remove the opener stepper motor.
ASSEMBLY:
(1)
Mount in reverse order. Observe IMPORTANT note below:
IMPORTANT:
Mount the motor with the motor cable outlet as
shown in the figure enclosed:
CALIBRATION:
(1)
Perform a stall calibration for the opener stepper motor.
Refer to chapter 3.6, Adjustments.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 129
Figure 130
Chapter 3.5 / 54