Checking And Adjusting The Height - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(2)
Shift the drawer unit upper plate
this way, that:
(3)
Fasten the lateral position
fixation screws.
2.1.3

Checking and Adjusting the Height

(1)
Insert the drawer unit adjustment tool in the cassette unit entry.
(2)
Check the minimum and maximum height. See also IMPORTANT note below.
IMPORTANT:
Perform the min. height check between
magnet and plastic finger.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
The plastic finger at the
drawer unit is aligned with
the lateral stop
While the drawer
adjustment tool is shifted to
the cassette input side wall.
Check
min.
height left.
Check
max.
height left.
max.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Adjustments and Calibrations
Lateral stop
517536an.cdr
Figure 10
Magnet
CR 30-X
Repair and Service
Plastic finger
Cassette input
side wall
517536az.cdr
Figure 9
Check
min.
height right.
min.
Check
max.
height right.
max.
Plastic Finger
min.
Figure 11
Chapter 3.6 / 9