Software Patch Cre_1206D; Solved Bugs With Software Cre_1206D Compared To Cre_1206C - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
4

Software Patch CRE_1206D

Type of new Software
Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
4.1

Solved Bugs with Software CRE_1206D compared to CRE_1206C

Solved insufficient initialization of the RAM on the PMI board, what caused interfering
noise in the image.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
Software Update
02/2007
 1366 plus 1187, 1260, 1261, 1263, 1338, 1355, 1357,
1363
MEDNET, GSO Library* (500 kB size)
Software CRE_1206 has to be installed before installing
the software patch CRE_1206D.
DD+DIS027.07E, Service Bulletin No. 06, Interfering
Noise in the Image:
Install Software Patch CRE_1206D (obsolete Document)
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Agfa Company Confidential
Chapter 3.8 / 8