DD+DIS150.06E
3.3
Checking the Shipment Completeness
(1)
Compare the scope of delivery with the packing list.
(2)
In case the delivery is not complete contact the Regional Support Network (RSN).
External partners: Contact your local Agfa Representative.
4
Performing Digitizer Setup
The purpose of this section is to show the required installation tasks.
The following table lists these tasks:
Task
Mount the distance holder.
Connect the cables.
Switch on the digitizer.
*Complete required time per Task Code to be entered in the Agfa repair report.
4.1
Mounting the Distance Holder
REQUIRED PARTS (part of delivery):
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
Required
Importance
Time
5 minutes
mandatory
2 minutes
mandatory
3 minutes
mandatory
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Task Code*
Section
INS
INS
INS
Figure 2
Chapter 1 / 8
Agfa Company Confidential
4.1
4.2
4.3