DD+DIS150.06E
(15) Open screws at input side of power supply.
(16) Remove power supply.
DANGER:
If the cables at the input side of the power
supply are mounted wrong, the housing of
the digitizer can be connected to high
voltage. Risk of live.
Mount the cables at the power supply input as
indicated on the cables and shown in Figure
250.
ASSEMBLY:
(1) Mount in reverse order.
See IMPORTANT note below.
IMPORTANT:
Connect the cables at the output side as shown
in Figure 251.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 249
Figure 250
black red
Figure 251
Chapter 3.5 / 103
Agfa Company Confidential