DD+DIS172.06E
Warning:
Images can be lost due to power failure.
Connect the equipment to an un-interruptible power supply (UPS) or an institutional
standby generator.
NOTE:
Previously a „CR Inroom X-Ray Shielding" was available as accessory.
This is not available anymore. Initiate appropriate measures to protect the digitizer
and cassettes with image plates from scattered radiation. See also section 7.1.
NOTE to Accessory "Remote Service":
In general it is possible to diagnose and repair the digitizer to a certain extent
remotely.
To enable remote service of the digitizer remote access has first to be organized.
Remote Access is established via the Agfa own Secure Remote Service System
(SRSS).
For more information to the SRSS see MedNet GSO Library path:
Technical Services > Service Delivery > SRSS
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2010
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Installation Planning
Chapter 11 / 10
Agfa Company Confidential