DD+DIS150.06E
5.2

Checking Software Version

NOTE
The digitizer only boots-up completely if the software version on the digitizer and the
software version on the control PC (Software for download to the digitizer) match. It
may happen that the delivered digitizer and the control PC do have a different
digitizer software version installed In this case error 2712 appears with each re-boot
of the control PC.
(1) Log-off the current user at the control PC.
(2) Login as user crservice.
(3) Select menu: <Start - Programs - Agfa CR 30-X - Service
Client>
(4) Enter user name and password.
(5) Select <Get installed version>.
(6) Evaluate as follows:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
Figure 12: Example screen after selecting "Get installed version"
If "Installed version" < "Version on PC":
Perform a software upload to the digitizer (Select nodes – Upload selected
nodes). For details refer to Service Documentation Chapter 3.2, Tools and
auxiliary
Means.
If "Installed version" > "Version on PC" (like in the example of Figure 12):
Perform digitizer software installation on the control PC.
The digitizer software is part of delivery.
Refer to the installation instructions on delivered software CD-ROM.
If "Installed version" = "Version on PC":
Okay. No further action required.