AGFA CR 30-X Service Manual page 266

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(5)
Insert a cassette the wrong way,
i.e. red side to the top and the
shutter inside the digitizer.
(6)
Push the cassette to the right: The
cassette edge sensor may not
switch:
(7)
If the sensor switches, open the
fixation screws at the cassette
edge sensor and push it more to
the right. Repeat step (5).
(8)
If the sensor does not switch,
repeat the complete procedure
once more, to confirm the proper
adjustment.
(1)
Switch on the digitizer.
Verification
(2)
Wait till the green light at the user interface lights up
(takes approximately 1 minute).
(3)
Insert a cassette the wrong way: The digitizer may not recognize the cassette.
(4)
Insert the cassette the right way: The digitizer has to recognize the cassette.
The cassette edge sensor is adjusted.
Result
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
No click may be audible.
When using a multimeter, the
resistance should be outside
the measuring range.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Adjustments and Calibrations
Wrong:
Red side on top
CR 30-X
Repair and Service
Figure 24
Chapter 3.6 / 18
Agfa Company Confidential