AGFA CR 30-X Service Manual page 168

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
(12) Take out cleaning brush holder.
(13) Remove screw at erasure unit.
(14) Disconnect two cables at erasure unit.
(15) Take out erasure unit.
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 56
Figure 57
Figure 58
Chapter 3.5 / 26
Agfa Company Confidential