Electrical Check after Power Supply Replacements
WARNING:
Improper ground connection of the digitizer metal housing may lead to electric
shocks.
In case of power supply replacements or maintenance check the protective earth
resistance according to the national regulations.
The electrical check is split up in two steps:
Step
Task
1
Perform the basic protective earth test.
2
Perform the electrical check according to National
Regulations.
3.3.5.1
Basic Protective Earth Test
(1) Use a multimeter that is capable to measure
a resistance < 1 Ohm
(2) Put one probe at the ground connector of
the mains input
(3) Put the other probe at a few metallic parts,
the customer could touch.
The electrical resistance has to be ≤ 0.2
Ohm.
Refer to Figure 12.