General Information on Preventive Maintenance

To ensure quality and functional reliability of the system all the points listed below
(minimum maintenance points) must be carried out.
IMPORTANT:
Check if it is necessary to include country specific regulations as additional
maintenance points.
NOTE:
Always have the latest digitizer software together with the associated NX workstation
software available on the Service PC. It can be downloaded from MEDNET, GSO
Library.
If the latest digitizer software version is not installed, consider a software upgrade of
the digitizer.
However always check, if it is compatible to the installed NX workstation software. It
may be required to upgrade the NX workstation software, too.
For details refer to section 2 of the digitizer service manual, appendix: Software /
Hardware Compatibility Chart.
1.1

Maintenance Frequency

The maintenance has to be carried out:
every 12.000 image plate cycles
or
once a year
1.2

Required Time

REQUIRED TIME:
approximately 2 hours
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009