Pmi Board Carrier Position Adjustment - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
2.2

PMI Board Carrier Position Adjustment

Adjust the position of the light collector this way that it is stress-free mounted.
Purpose
Miss-alignment can lead to IP jams, laser reflections or cracks in the light collector.
After replacement of the Photomultiplier (PMT) with light collector
Use Case
REQUIRED TOOLS:
REQUIRED TIME:
30 minutes
IMPORTANT:
The PMI Board Carrier Position Adjustment is part of the replacement procedure "PMT
with Light Collector".
The adjustment has to be done before mounting the PMT, after mounting optic module
and PMI board carrier.
MECHANICAL ADJUSTMENT:
(1)
Shift the PMI carrier to the rear.
(2)
Fasten the four fixation screws.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
13 mm
PMT gauge
(Part of delivery)
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Adjustments and Calibrations
6 mm
Figure 15
Agfa Company Confidential
Chapter 3.6 / 12

Hide quick links: