DD+DIS150.06E
2.2.7

Service Client Menu "SW Upload"

The "SW Upload" menu allows:
Purpose
To check if the software version currently installed on the control PC
matches with this installed on the digitizer.
To upload the digitizer software to a specific single, two or all 3 PCB's (nodes)
in the digitizer.
To upload the software from the control PC to the digitizer may be required in the
following cases:
After replacement of the PMI board, slow scan board or IP handling board
To perform a software update (bug-fixing) or SW upgrade (new functionality)
NOTE:
All steps for software upgrade / update of the digitizer are described in a document
which is part of the software on MedNet GSO Library (readme or enclosure document).
Depending on the software version, the procedure might be different to the steps
described here.
(1)
Select "General Device Info - SW Versioning".
Usage
(2)
Select "Get installed version". See Figure 14.
Digitizer
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Node
Installed Version Version on PC
CRE_1206
PMI Board
CRE_1206
HDL Board
CRE_1206
SSC Board
Get installed version
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Tools and auxiliary Means
Status
ok
CRE_1206
ok
CRE_1206
ok
CRE_1206
Upload selected nodes