DD+DIS150.06E
(5)
Place the cassette in length
direction to the X-ray tube.
(6)
Ensure that the entire image plate
is fully exposed! The collimated
field must be larger than the image
plate.
(7)
Select the following exposure parameters to obtain a dose of 10 µGy:
(8)
Expose the cassette.
(9)
Turn the cassette 180°.
(10) Expose the cassette again.
After the second exposure the cassette has been exposed in total
with a dose of approximately 20 µGy.
(11) Remove the Cu-filter from the X-ray device.
(12) Start the CR 30-X Service Client: Select "Start - Programs - Agfa CR 30-X -
Service Client".
(13) Select "Preventive Maintenance - Calibration":
The current shading calibration is displayed. See Figure 43.
(14) Select "Calibrate".
(15) Insert the exposed cassette: The IP is scanned and erased.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
12 mAs
75 kV
1.3 m (51,2") distance
Large focus
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Adjustments and Calibrations
Rotating
Anode
OK
517536ab.cdr
OK
CR 30-X
Repair and Service
X-Ray tube
517536aa.cdr
NOT OK
Figure 41
Exposed
area
Cassette
NOT OK
Figure 42
Chapter 3.6 / 30