(28) Take off the complete lower
transport unit.
(29) Remove 3 screws at each brush.
(30) Take out the brushes.
ASSEMBLY:
(1)
Mount in reverse order.
(1) Switch on the digitizer.
Verification:
(2) Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3) Insert cassette with unexposed IP in the digitizer.
(4) At the workstation select examination type "System diagnosis – Flatfield".
(5) Select "ID" to identify the image: Scanning starts.
(6) Confirm that the image arrives at the workstation.
(7) Reject the image.
The Antistatic Brush Cassette Unit Entry is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 40
Figure 41
Chapter 3.5 / 19