2.1.1

Troubleshooting Software Installation Problems at the Control PC

The following problems may occur during software installation at the control PC:
After de-installation of the old digitizer software on the control PC following error
Symptom 1
message pops-up:
The NX software or the error viewer have not been closed before software de-
Cause
installation.
Restart the control PC. Continue with software installation.
Solution
See step (9) previous page.
After reboot of the control PC the screen "Welcome to the Found New Hardware
Symptom 2
Wizard" appears.