DD+DIS150.06E
2.2.2
Service Client Menu "E-Label"
The 'E-labels' (used in 5 different modules, see Figure 9) keep device and module
Purpose
specific data, like module serial number, last error codes or module specific
parameters.
Purpose of the E-Labels:
Keep module specific data to adjust settings in the digitizer automatically.
Ensure automatic tracking of module exchange (automatic infocounter entries).
Allow fault investigations in production, as the returned, defective module has all
required data in the E-Label.
(1) Select "General Device Info - E-Label".
Usage
(2) Select the desired E-Label to view the contents.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Figure 9
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 13