DD+DIS150.06E
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
(12) Remove screw at erasure unit.
(13) Disconnect two cables at erasure unit.
(14) Take out erasure unit.
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(15) Remove IP guide plate.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 188