AGFA CR 30-X Service Manual page 238

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(9)
Remove 2 screws at safety switch left and
right hand side.
(10) Remove safety switch.
(11) Remove cables at safety switch.
ASSEMBLY:
(1)
Mount in reverse order.
MECHANICAL ADJUSTMENT:
(1)
Perform the cassette unit safety switch position adjustment.
Refer to section 3.6, adjustments.
(1)
Switch on the digitizer.
Verification:
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The Cassette Unit open Safety Switch is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 233
Chapter 3.5 / 96
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents