Maintenance - Power Off; Visual Inspection - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS171.06E
3
Maintenance – Power Off
3.1

Visual Inspection

REQUIRED TOOLS:
Flashlight
Hex key 3 mm
(1)
Switch off the digitizer.
(2)
Unplug the mains cable at the digitizer.
(3)
Check overall condition of the device for obvious changes or damage.
(4)
Check the condition of the power and FireWire cable.
(5)
Open cassette unit.
(6)
Remove top cover.
(7)
Remove front cover.
(8)
Check the condition of the cables
inside the digitizer, especially in the
area of the cassette unit.
(9)
If a wear of the cassette unit wire
harness can be observed (see
arrow):
Replace the side plate by "Side Plate
Transport Unit (left hand side)".
Spare part number see section 1.6.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
Cassette side plate with straight edge
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Figure 8: Removing the top cover
Figure 9:
(old design)
Agfa Company Confidential