AGFA CR 30-X Service Manual page 51

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(2)
Insert an edgeless means, e.g. a
pen into the cassette lock.
(3)
Open the shutter.
(4)
Turn the cassette around, so that
the black tube side is above.
(5)
Let the black tray and image plate
slide out carefully onto the table.
(6)
Put the black tray with the image
plate onto the cassette.
(7)
Check the image plate for
scratches.
(8)
Clean the image plate if required.
(9)
Put the tray with the image plate
back into the cassette.
(10) Verify that the white phosphor side
is oriented to the black tube side of
the cassette and that the shutter
does not scratch the image plate.
(11) Insert an edgeless means, e.g. a
pen into the cassette lock while
closing the shutter.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
517509ag.cdr
Figure 14
1
2
1
517509ad.cdr
Figure 15
Figure 16
White phosphor side
1
2
Black tube side of the cassette
517509af.cdr
Figure 17
Agfa Company Confidential
Chapter 1 / 17