Functional Description
4
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Chapter 2 / 6
Agfa Company Confidential

Digitizer Boot-Up

Digitizer
Switch-On
Boot-up of PMI
and all other CAN nodes
NX connected?
2
YES
3
Send Self Test
Signal to all SW modules
Selftest okay?
YES
IP Clean-up
Cycle
Cassette or IP
found?
YES
NO
Initialize all Modules
= set all motors to HOME
Initialization okay?
YES
NX Viewer
running?
YES
Digitizer idle
Legend:
1
3 x beep indicates: PMI + CAN nodes are started up.
Connection only possible if
2
- FireWire cable connected
- ASAP service on NX started.
DOCUMENT CONTROL NOTE:
Type 5175/100
CR30-X Boot-Up
repeat endless
if required
NO
Error message
in NX Error Viewer.
NO
Boot-up stops
Warning message
in Error Viewer
Remove Cassette
or IP and continue
Error message
in NX Error Viewer.
NO
Boot-up stops
Wait till the NX
NO
sends "enable feeder"
message (LED blinks red)
3
3
This selftest is without motor actvities (except polygon).
It checks e.g. whether motors and sensors are connected.
Figure 6
CR30-X
DD+DIS150.06E
Audible
3 x
1
beep
SW Modules:
- Image Capture Control
- Cassette Handling Control
- IP Handling Control
- Slow Scan Control
Edition 1, Revision 0
Visible
517502eb.cdr

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents