DD+DIS171.06E
3.7
Cleaning of the lower Scan Rollers
IMPORTANT:
Spot artifacts may be caused by a dirty scan roller, which transfers the dirt particles
during scan to the image plate.
Therefore, only clean the scan rollers, if sport artifacts are visible in the image made
under step 2.6.
Otherwise skip this step and re-mount all covers.
REQUIRED TOOLS:
•
Lint free cloth
•
ADC screen cleaner
•
Hex key 3 mm and 5 mm
IMPORTANT:
Do not clean the upper rollers: They do not require any cleaning.
(1)
Remove the left and right side
covers.
(2)
Remove the rear cover.
(3)
Remove the guide plate before the
pre-scan rollers (2 screws left and
right see arrows).
(4)
Clean the lower pre-scan roller with
a lint free cloth lightly soaked with
CR screen cleaner.
Turn the clutch at the slow scan unit
to turn the rollers.
Pushing the clutch approx. 60 times
gives one roller rotation.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Figure 21
Chapter 9 / 20
Agfa Company Confidential