AGFA CR 30-X Service Manual page 241

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(13) Disconnect cables at Distributor board.
(14) Remove distributor board.
ASSEMBLY:
(1) Mount in reverse order.
(1)
Switch on the digitizer.
Verification:
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The Distributor Board is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.