3.2

Cleaning General

REQUIRED TOOLS:
Vacuum cleaner
(1)
Clean the inside of the digitizer with
a vacuum cleaner.
Especially check for loose dust in
transport unit, and clean this
thoroughly with the vacuum cleaner.
3.3

Cleaning of the Erasure Unit

REQUIRED TOOLS:
Lint free cloth
(1)
Check the reflector, i.e. the lower
side of the erasure unit.
Remove dirt or dust with a lint free
cloth.