DD+DIS150.06E
2.2.10

Service Client Menu "Endurance Run Cycle"

The "Endurance Run Cycle" allows to run any amount of cycles with/ without scanning
Purpose
and/ or erasure. This can be helpful when troubleshooting sporadic image plate
handling failures.
(1) Select "Analysis & Repair - Endurance run Cycle".
Usage
(2) Put a cassette into the digitizer.
(3) Select the number of cycles and delay between cycles (if required).
(4) Select "Scanning" and/or "Erasure" if required.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.