Laser Beam Position Check And Adjustment - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
2.5

Laser Beam Position Check and Adjustment

The purpose of the "Laser Beam Position Check and Adjustment" is:
Purpose
Checking the laser beam position, to see, whether it is centered to the calibration
board.
Adjustment of the optic module this way that the laser beam position is centered
to the calibration board.
Perform the "Laser Beam Position Check and Adjustment" in following cases:
Use Case
In case of sporadic or persistent errors 10761 or 10767 to identify the possible
root cause of an incorrect laser beam position
To solve errors 10761 or 10767, if an incorrect laser beam position is the root
cause
REQUIRED TOOLS:
REQUIRED TIME:
Approximately 30 minutes
IMPORTANT:
This adjustment requires:
Calibration board version 2* and
Distributor board version 2* (see Figure 25) and
Software ≥ CRE_1511**
For digitizers with lower serial numbers, which do not fulfil the requirements, it can also
be performed with the Laser Beam Detection Toolkit, spare part number***:
CM+9 5175 2940 0
The Laser Beam Detection Toolkit has instructions for installation and adjustment
included.
* Installed in SN ≥ 6397 (Type 5175/100) and SN ≥ 5935 (Type 5175/110) and in all digitizers, where the
calibration board was exchanged after September 2009.
** Installed in SN ≥ 5570 (Type 5175/100) and SN ≥ 6377 (Type 5175/110).
*** The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
3 mm
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Adjustments and Calibrations
Agfa Company Confidential